Command reference
R&S
®
ZNA
1577User Manual 1178.6462.02 ─ 29
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe <Boolean>
Enables or disables the source power calibration for channel <Ch> and for port number
<PhyPt>. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (SOURce<Ch>:POWer:CORRection[:ACQuire]).
The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe].
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire].
Manual operation: See "Port Overview" on page 637
SOURce<Ch>:POWer<PhyPt>:CORRection:UPORt:STATe <Boolean>
This command can only be used on a 2-port R&S ZNAxx with Chapter 4.7.23, "Internal
2nd source and 2nd LO generator for 2-port R&S ZNA", on page 303 and internal
combiner.
If the internal combiner is used (SOURce<Ch>:COMBiner INTernal),
SOURce<Ch>:POWer2:CORRection:UPORt:STATe ON|OFF enables or disables
the source flatness cal of the upper-tone port 2, previously performed using
SOURce<Ch>:POWer:CORRection[:ACQuire] UPORt, 2.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number. Must be 2.
Parameters:
<Boolean> ON (1) or OFF (0)
Manual operation: See "Port Overview" on page 597
7.3.15.11 SOURce:POWer:CPRemeas commands
Theses commands enable and confure the calibration premeasurement feature (see
Chapter 4.1.6, "Power correction from pre-measurement", on page 119).
SOURce<Ch>:POWer<Pt>:CPRemeas:CFACtor <Value>
Specifies the convergence factor for a cal premeas at port <Pt> in channel <Ch>.
SCPI command reference