Parameters:
<Boolean> ON use independent segment sweep time settings, to be
defined via [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME.
OFF reset the segment sweep time in all sweep segments to
the segment sweep time for unsegmented sweeps, defined via
[SENSe<Ch>:]SWEep:TIME. ON will not restore the previous
values.
The parameter is automatically switched to ON when a segment
sweep time is entered using [SENSe<Ch>:]SEGMent<Seg>:
SWEep:TIME or if the channel settings in a sweep segment
require a sweep time larger than the unsegmented sweep time.
*RST: OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1.
SEGM:SWE:TIME 0.1
Increase the segment sweep time to 0.1 s.
SEGM:SWE:TIME:CONT OFF
Couple the sweep times in all segments and reset the sweep
time in segment no. 1 to the initial value.
Manual operation: See "Optional Columns" on page 557
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?
Returns the total duration of the segmented sweep, calculated as the sum of the
sweep times of the individual segments ([SENSe<Ch>:]SEGMent<Seg>:SWEep:
TIME).
Suffix:
<Ch>
.
Channel number.
<Seg> Sweep segment number. This suffix is ignored; the command
returns the sum of all segments.
Example: See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
Usage: Query only
Manual operation: See "Optional Columns" on page 557
7.3.14.21 [SENSe:]SPECtrum...
[SENSe<Ch>:]SPECtrum:DETector............................................................................... 1497
[SENSe<Ch>:]SPECtrum:IREJection............................................................................. 1498
[SENSe<Ch>:]SPECtrum:ZPADing................................................................................ 1498
[SENSe<Ch>:]SPECtrum:DETector <Type>
Defines the detector for the spectrum measurement, i.e. how the measurement result
for a particular sweep point is derived from the related sample values.