Command reference
R&S
®
ZNA
1514User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]SWEep:TIME <SweepDuration>
Sets the duration of the sweep ("Sweep Time"). Setting a sweep time disables the
automatic calculation of the (minimum) sweep time; see [SENSe<Ch>:]SWEep:
TIME:AUTO.
Note: The sweep time is ignored for the sweep types "Time" and "CW Mode"
([SENSe<Ch>:]SWEep:TYPE).
Suffix:
<Ch>
.
Channel number
Parameters:
<SweepDuration> Sweep time. The minimum possible sweep time depends on the
other channel settings, in particular on the number of points
([SENSe<Ch>:]SWEep:POINts), the IF bandwidth
([SENSe<Ch>:]BWIDth[:RESolution]) and the measure-
ment delay for each partial measurement ([SENSe<Ch>:
]SWEep:DWELl). Changing the duration leaves the number of
points unchanged but directly affects the delay.
Range: Between minimum sweep time (for current channel
settings) and 100000 s
*RST: Minimum sweep time (for *RST channel settings)
Default unit: s
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:DWEL?
Query the delay for each partial measurement.
SWE:TIME 2
Increase the total sweep time to 2 s.
SWE:DWEL?
Query the meas. delay for each partial measurement again. The
delay is increased by 1 s divided by the total number of partial
measurements per sweep.
Manual operation: See "Stop Time" on page 524
[SENSe<Ch>:]SWEep:TIME:AUTO <Boolean>
When enabled, the (minimum) sweep time is calculated internally using the other chan-
nel settings and zero delay ([SENSe<Ch>:]SWEep:DWELl).
Note: The automatically calculated sweep duration is ignored for the sweep types
"Time" and "CW Mode" ([SENSe<Ch>:]SWEep:TYPE).
Suffix:
<Ch>
.
Channel number
SCPI command reference

Table of Contents

Related product manuals