GUI reference
R&S
®
ZNA
627User Manual 1178.6462.02 ─ 29
Efficient power calibration procedure
●
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
●
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
●
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to Chapter 4.5.6, "Scalar power calibration", on page 218.
5.11.3.1 Controls on the Pwr Cal Settings tab
Some buttons serve as openers for additional dialogs:
●
"Cal Power...": see Chapter 5.11.3.2, "Cal Power Config dialog", on page 629
●
"Transm. Coefficients...": see Chapter 5.11.3.3, "Power Meter Transmission Coeffi-
cients dialog", on page 632
●
"Power Meters...": see Chapter 4.7.39, "External power meters", on page 316
Cal softtool