GUI reference
R&S
®
ZNA
627User Manual 1178.6462.02 ─ 29
Efficient power calibration procedure
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to Chapter 4.5.6, "Scalar power calibration", on page 218.
5.11.3.1 Controls on the Pwr Cal Settings tab
Some buttons serve as openers for additional dialogs:
"Cal Power...": see Chapter 5.11.3.2, "Cal Power Config dialog", on page 629
"Transm. Coefficients...": see Chapter 5.11.3.3, "Power Meter Transmission Coeffi-
cients dialog", on page 632
"Power Meters...": see Chapter 4.7.39, "External power meters", on page 316
Cal softtool