EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 5.12.1.2 Channel Manager dialog

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
655User Manual 1178.6462.02 ─ 29
Delete Channel
Deletes the current channel including all the traces assigned to it and removes all dis-
play elements related to the channel from the diagram area. "Delete Channel" is dis-
abled if the recall set contains only one channel: In manual control, each recall set
must contain at least one diagram area with one channel and one trace.
Tips: Use the "Undo" function to restore a channel that was unintentionally deleted.
Remote command:
CONFigure:CHANnel<Ch>[:STATe] OFF
Channel On
Toggles the measurement state of the active channel.
Remote command:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Only Active Channel On
If checked, only the active channel is measured; all other channels are switched off
temporarily and dynamically.
Temporarily means: If "Only Active Channel On" is unchecked, then all channels return
to their original Channel On state.
Dynamically means: If another channel becomes the active one, the firmware adjusts
the temporary ON/OFF state accordingly.
Remote command:
CONFigure:CHANnel:MEASure:ACTive[:STATe]
Fixture Simulator
The "Fixture Simulator" switch deactivates or activates the configured deembedding,
embedding, balanced ports, and port impedance settings for the selected channel.
When "Fixture Simulator" is deactivated:
All balanced ports are resolved to single ended ports
All port impedances are set to default
All de/embeddings are disabled
At the GUI, the "Balanced Ports" dialog and the de/embedding tabs and dock widgets
are disabled. Related remote commands generate an error.
When "Fixture Simulator" is reactivated, the situation before the deactivation is
restored.
Note: The "Offset" and "One Way Loss" settings are not affected by the "Fixture Simu-
lator" switch. Use the "All Compensation Off"/"All Compensation On" functions of the
"Fixture Compensation" dialog to de/activate these compensations as well (see Chap-
ter 5.14.2.2, "Fixture Compensation dialog", on page 719).
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:FSIMulator[:STATe]
5.12.1.2 Channel Manager dialog
The "Channel Manager" dialog allows you to rename channels and to change their
measurement state and sweep mode.
Channel Config softtool

Table of Contents

Related product manuals