1203User Manual 1178.6462.02 ─ 29
CONFigure:CHANnel:CATalog?
Returns the numbers and names of all channels in the current recall set. The response
is a string containing a comma-separated list of channel numbers and names; see
example below. If all channels have been deleted the response is an empty string ("").
Example:
*RST; :CONF:CHAN2:STAT ON; NAME 'New Channel'
Create channel 2 and assign the channel name "New Channel".
CONF:CHAN:CAT?
Query all channels and their names. As a default channel no. 1
is created on *RST, the response is
'1,Ch1,2,New_Channel'.
CONF:CHAN:NAME:ID? 'New Channel'
Query the channel number for the channel named "New Chan-
nel". The response is 2.
Usage: Query only
Manual operation: See "Channel table" on page 656
CONFigure:CHANnel:MEASure:ACTive[:STATe] <Boolean>
If set to ON (1), only the active channel is measured. All other channels are switched
off temporarily, i.e. until CONFigure:CHANnel:MEASure:ACTive is set to OFF.
Parameters:
<Boolean>
ON (1) | OFF (0)
Manual operation: See "Only Active Channel On" on page 655
CONFigure:CHANnel:MEASure:ALL[:STATe] <Boolean>
Enables or disables the sweep in all channels of the active recall set. This command
can be used in combination with CONFigure:CHANnel<Ch>:MEASure[:STATe] to
optimize the measurement speed.
Parameters:
<Boolean> ON | OFF
*RST: ON
Example: See CONFigure:CHANnel<Ch>:MEASure[:STATe]
Manual operation: See "Continuous / Single" on page 565
CONFigure:CHANnel:MEASure:OPTimized <Order>[, <Ch1>, <Ch2>, ...]
This setting tells the firmware to optimize the channel switching times for a particular
channel sweep order.