Command reference
R&S
®
ZNA
1119User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - fail beep on or off.
*RST: OFF
Example:
CALC:RIPP:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation: See "Ripple Fail Beep" on page 489
CALCulate<Chn>:RIPPle:STATe <Boolean>
Switches the ripple limit check for the active trace on or off.
Tip: Use CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe] to switch the limit
check for a single ripple limit range on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF – ripple limit check on or off.
*RST: OFF
Example: See CALCulate<Chn>:RIPPle:FAIL?
Manual operation: See "Ripple Check" on page 488
CALCulate<Chn>:RIPPle:STATe:AREA <HorizontalPos>, <VerticalPos>
Moves the ripple test info field for the active trace <Chn> to one of nine predefined
positions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:RIPPle:FAIL?
Manual operation: See "Ripple Check" on page 488
7.3.1.13 CALCulate:STATistics...
The CALCulate:STATistics... commands evaluate and display statistical and
phase information of the trace.
SCPI command reference