Command reference
R&S
®
ZNA
1114User Manual 1178.6462.02 ─ 29
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus
range between 2 GHz and 3 GHz, assigning a ripple limit of +5
dB.
CALC:RIPP:DEL:ALL
Delete both created ripple limit ranges.
Usage: Event
Manual operation: See "Add / Insert / Delete / Delete All / Align All" on page 491
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe] <Enable>
Configures the display of ripple check info fields for the active recall set.
Parameters:
<Enable> ON - Info fields are displayed for all traces, for which a limit
check is enabled.
OFF - Only the info field for the active trace is displayed (if the
ripple check is enabled for this trace).
*RST: OFF
Manual operation: See "Show Results All Traces" on page 489
CALCulate<Chn>:RIPPle:DISPlay[:STATe] <Boolean>
Displays or hides all ripple limit lines (including all ranges) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - ripple limit line on or off.
*RST: OFF
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:DISP ON
Show the ripple limit range in the active diagram.
Manual operation: See "Show Ripple Limits" on page 488
CALCulate<Chn>:RIPPle:FAIL?
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed.