GUI reference
R&S
®
ZNA
488User Manual 1178.6462.02 ─ 29
Show Ripple Limits
Shows or hides the ripple limit lines associated with the active trace in a Cartesian dia-
gram area. The vertical positions of the ripple lines are recalculated after each sweep;
only their stimulus range and distance (the ripple limit) are fixed.
Note: Limit line display and limit check are independent of each other: Hiding the limit
line does not switch off the limit check.
Remote command:
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
Ripple Check
Switches the ripple limit check of the active trace on or off. When the limit check is
switched on, a movable info field shows the pass/fail information and the measured rip-
ple in each ripple limit range.
Limit violations are marked with squares, whose color is defined in the Define User
Color Scheme dialog ("Limit Fail Trace Color").
An acoustic signal and a TTL signal indicating pass or fail can be generated in addi-
tion.