EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - CALCulate:TRANsform:VNETworks...

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1144User Manual 1178.6462.02 ─ 29
Parameters:
<Unit> TIME | DISTance
TIME - x-axis scaled in time units.
DISTance - x-axis scaled in distance units (Distance = Time * c
0
* Velocity Factor).
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:XAX DIST
Convert the x-axis scaling to distance units.
Manual operation: See " Time / Distance" on page 535
CALCulate:TRANsform:VNETworks...
The CALCulate:TRANsform:VNETworks... commands define and activate the cir-
cuit models for the different de-/embedding types. SeeChapter 4.6.2, "Embedding and
deembedding", on page 234 .
The circuit models are addressed by predefined enum constants. Except for the FIM-
Port model, that consists of a single snp file, the models and constants are different for
the different de-/embedding types.
Table 7-6: Circuit models for single ended port embedding/deembedding
Parameter Circuit model Pictogram
FIMPort File import, generic 2-port (no circuit model)
CSL Serial C, shunt L
LSC Serial L, shunt C
CSC Serial C, shunt C
SCPI command reference

Table of Contents

Related product manuals