Receiver Level tab
Provides access to the receiver step attenuator settings.
Receiver step attenuators are optional hardware (see Chapter 4.7.31, "Receiver step
attenuators", on page 309).
Attenuation
Sets the attenuation at the respective measurement receiver (b-wave). See "Receiver
Step Att." on page 527.
Remote command:
[SENSe<Ch>:]POWer:ATTenuation
Wb/Nb IF Gain
Controls the gain of the wideband and narrowband signal paths.
Same functionality as "IF Gain Mode" on the Mode tab.
Remote command:
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
Input/Output tab
This tab gives access the settings for the optional Direct generator/receiver access and
Chapter 4.7.30, "Direct IF access", on page 308.
If a port is configured as converter port, all its "Input / Output" settings are derived from
the settings in the Converter Configuration dialog. Hence the corresponding table row
is read-only.