EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 678

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
678User Manual 1178.6462.02 ─ 29
Background information
Refer to Chapter 4.7.27, "Direct generator/receiver access", on page 306 and Chap-
ter 4.7.30, "Direct IF access", on page 308.
Receiver Input
Determines how the MEAS and REF signals are routed to the respective receivers.
Depending on the selected path, the hardware and factory correction values are set. If
"Port" is used, also the factory calibration is applied. This setting ensures reasonably
corrected data with the different cabling options.
"Port"
Via the standard port and with standard jumpering of the Direct gen-
erator/receiver access connectors.
"Direct
Access"
Via the direct generator/receiver access connectors (see Chap-
ter 4.7.27, "Direct generator/receiver access", on page 306).
"Reverse Cou-
pler"
Via the standard port but with reversed coupler configuration (see
"Reverse Coupler Configuration" on page 307).
"Source Moni-
tor"
Via the direct source monitor access connector (see Chapter 4.7.34,
"Direct source monitor access", on page 310).
Can only be selected for analyzer ports 1 and 3, and only if the
optional source monitor access is available for this port.
"Rear IN"
Via the direct IF access connectors (see Chapter 4.7.30, "Direct IF
access", on page 308).
For "Direct Access", "Reverse Coupler", and "Source Monitor" access, the LED above
the direct generator/receiver access connectors is switched on
For analyzer ports that are used as converter ports, the "Receiver Input" is configured
in the Converter Configuration dialog.
Remote command:
[SENSe<Ch>:]PATH<Pt>:DIRectaccess
SOURce<Ch>:PATH<Pt>:DIRectaccess
[SENSe<Ch>:]PATH<Pt>:MEASurement:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:REFerence:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
Rear Out
If checked, the Direct IF access connectors on the rear panel are used as (additional)
output.
Remote command:
[SENSe<Ch>:]PATH<Pt>:IFSWitch
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
Auto Determine IF Out
This checkbox is only enabled, if Rear Out is used.
Channel Config softtool