1462User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFOutauto?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFOutauto?
Queries how the outgoing frequencies at IF Reference <Pt> and IF Meas <Pt> are
determined in channel <Ch> if [SENSe<Ch>:]PATH<Pt>:IFSWitch is set to OUT.
To define how the frequencies are determined, use [SENSe<Ch>:]PATH<Pt>:
IFOutauto.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Auto Determine IF Out" on page 678
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFRequency?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFRequency?
Queries the freqency conversion formula for Direct IF Access Input / Output ports IF
Reference <Pt> and IF Meas <Pt> in channel <Ch>.
To set the formula, use [SENSe<Ch>:]PATH<Pt>:IFRequency
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Rear IF Frequency" on page 679
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
Queries if and how the Direct IF Access Input / Output ports IF Reference <Pt> and IF
Meas <Pt> are used in channel <Ch>.
To define the port usage, use [SENSe<Ch>:]PATH<Pt>:IFSWitch.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Rear Out" on page 678