Command reference
R&S
®
ZNA
1459User Manual 1178.6462.02 ─ 29
LP_60m
Low pass, frequency range 5 kHz to 60 MHz
BP_279m
Band pass, frequency range 260 MHz to 290 MHz
Usage: Setting only
Options: R&S ZNA-B26
Manual operation: See "Rear Input Path" on page 679
[SENSe<Ch>:]PATH<Pt>:IFOutauto
<Boolean>
Defines how the outgoing frequencies at IF Reference <Pt> and IF Meas <Pt> are
determined in channel <Ch>.
This setting only takes effect if [SENSe<Ch>:]PATH<Pt>:IFSWitch is set to OUT.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Setting parameters:
<Boolean> OFF
The frequencies are set manually.
Use [SENSe<Ch>:]PATH<Pt>:IFRequency to define a suita-
ble frequency conversion formula.
ON
The frequencies are set automatically (by the analyzer firm-
ware).
*RST: OFF
Usage: Setting only
Options: R&S ZNA-B26
Manual operation: See "Auto Determine IF Out" on page 678
[SENSe<Ch>:]PATH<Pt>:IFRequency <Numerator>, <Denominator>, <Offset>,
<Sweep type>
Defines the freqency conversion formula for Direct IF Access Input / Output ports IF
Reference <Pt> and IF Meas <Pt> in channel <Ch>.
For <Sweep type> FB, the IF frequency f
IF
is calculated according to
f
IF
= <Numerator>/<Denominator> * f
b
+ <Offset>,
where f
b
represents channel base frequency.
For <Sweep Type> CW, f
IF
is calculated according to
f
IF
= <Offset>,
SCPI command reference