GUI reference
R&S
®
ZNA
455User Manual 1178.6462.02 ─ 29
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
Time Gate Linking/Single Time Gate Linking
Allows you to link the time gates of a channel's time domain traces, i.e. to align the
time gate setting across linked (~coupled) traces.
"Time Gate Linking" links the time gates of all traces to the time gate of the active
trace, "Single Time Gate Linking" links the time gate of the current trace to the time
gate of the selected trace.
The controls in these sections are enabled, if the current channel contains at least one
time domain trace,
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:LINK
5.5.7 Distance to Fault tab
This tab is only visible, if option R&S ZNA-K2 is installed on the R&S ZNA. It replicates
the "Distance to Fault" tab of the "Distance to Fault" application. See Chapter 5.17.3,
"Distance to Fault application", on page 830.
5.5.8 Trace Statistics tab
Evaluates statistical and phase information of the entire trace or of a specific evalua-
tion range and calculates the x-dB compression point.
Traces softtool