GUI reference
R&S
®
ZNA
742User Manual 1178.6462.02 ─ 29
DC Extrapolation
If unchecked (default), the touchstone data of the generated deembedding file reach
down to the start frequency of the instrument sweep. If checked, the EZD tool adds
(extrapolated) S-parameter data for f = 0.
Remote command:
CALCulate:FMODel:EZD:DCEXtrapolat
Impedance Correction Settings
Provides additional settings for the impedance correction of the EZD tool, if enabled in
the main dialog (see Use Impedance Correction).
Remote command:
CALCulate:FMODel:EZD:IMPedance:REFerence
CALCulate:FMODel:EZD:IMPedance:BWIDlimit
Reset to Default
Resets the settings of the "EZD Advanced Settings" to their defaults.
Remote command:
CALCulate:FMODel:EZD:PRESet
5.14.6 Port Sets tab
Selects transformation networks for deembedding/embedding arbitrary port sets,
defines their parameters, assigns them to a port set and enables embedding.
Use the complementary dock widget to create the required port sets and to activate or
deactivate dembedding/embedding for selected port sets (see Chapter 5.14.1.4, "Port
Sets panel", on page 710).
Background information
Refer to Chapter 4.6.2.5, "Port pair de-/embedding", on page 238 and Chapter 4.6.2.6,
"Port set de-/embedding", on page 239.
Offset Embed softtool