EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 5.14.6 Port Sets tab

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
742User Manual 1178.6462.02 ─ 29
DC Extrapolation
If unchecked (default), the touchstone data of the generated deembedding file reach
down to the start frequency of the instrument sweep. If checked, the EZD tool adds
(extrapolated) S-parameter data for f = 0.
Remote command:
CALCulate:FMODel:EZD:DCEXtrapolat
Impedance Correction Settings
Provides additional settings for the impedance correction of the EZD tool, if enabled in
the main dialog (see Use Impedance Correction).
Remote command:
CALCulate:FMODel:EZD:IMPedance:REFerence
CALCulate:FMODel:EZD:IMPedance:BWIDlimit
Reset to Default
Resets the settings of the "EZD Advanced Settings" to their defaults.
Remote command:
CALCulate:FMODel:EZD:PRESet
5.14.6 Port Sets tab
Selects transformation networks for deembedding/embedding arbitrary port sets,
defines their parameters, assigns them to a port set and enables embedding.
Use the complementary dock widget to create the required port sets and to activate or
deactivate dembedding/embedding for selected port sets (see Chapter 5.14.1.4, "Port
Sets panel", on page 710).
Background information
Refer to Chapter 4.6.2.5, "Port pair de-/embedding", on page 238 and Chapter 4.6.2.6,
"Port set de-/embedding", on page 239.
Offset Embed softtool

Table of Contents

Related product manuals