GUI reference
R&S
®
ZNA
743User Manual 1178.6462.02 ─ 29
If the "Fixture Simulator" is disabled for the related channel (see "Fixture Simulator"
on page 655), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See "Overview" on page 717.
Type
Switches between "Deembedding" and "Embedding" network definition.
Port Set
Port sets, defined in the complementary Port Sets panel dock widget panel. The trans-
formation networks are defined such that the physical analyzer test ports are connec-
ted to the left of the circuit; the DUT ports are on the right side. You can define inde-
pendent transformation networks for all port sets.
The port set number, i.e. the number at the beginning of each "Port Set" item, corre-
sponds to the position of the port set in the Port Sets panel.
Remote command:
The port set number corresponds to the <ListId> numeric suffix in the port set de-/
embedding commands; see e.g. CALCulate<Ch>:TRANsform:VNETworks:PPAir:
DEEMbedding<ListId>[:STATe].
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if Type = "Deembedding".
Network
Depending on the size m of the selected port set, the graphical list contains all availa-
ble 2m-port deembedding/embedding networks.
Offset Embed softtool