GUI reference
R&S
®
ZNA
599User Manual 1178.6462.02 ─ 29
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
SOURce<Ch>:POWer<PhyPt>:CORRection:UPORt:STATe on page 1577
Apply
Is enabled when a new set of power calibration data has been acquired. The button
applies all available power calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
The power calibration state is indicated in the trace list, see Chapter 4.5.6.3, "Power
calibration labels", on page 222. Use the functions in the Chapter 5.11.4, "Use Cal tab",
on page 634 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>
[SENSe<Chn>:]CORRection:PSTate?
Power Cal dialog – Ref. Receiver
In Ref. Receiver mode, the "Power Cal" dialog guides you through a reference receiver
calibration.
The selected source port is displayed with its cal power settings (see Chapter 5.11.3.2,
"Cal Power Config dialog", on page 629). A circuit diagram visualizes the connectivity.
A power meter must be connected to the calibrated source port.
Access: Power Cal wizard > Ref. Receiver
Background information
Refer to Chapter 4.5.6.1, "Source power calibration", on page 220.
Cal softtool