GUI reference
R&S
®
ZNA
598User Manual 1178.6462.02 ─ 29
Only a source flatness calibration can be performed, and this calibration can only
be done using a power meter (irrespective of the selected Power Cal Method.
If option R&S ZNA is available and frequency converters are configured, additional
source flatness calibrations can be performed for the analyzer ports driving the
converters' RF IN and LO IN ports.
In the converter configuration above, P1 and P2 drive the RF IN ports of converter
1 and 2, respectively, and the Converter LO port drives the LO IN ports of both con-
verters.
Since firmware version 2.40, the "Leveling Table" buttons give access to the con-
verter leveling functionality that previously required a separate software.
Select a "Ref. Receiver" or "Meas. Receiver", "Leveling Table", or "Source Flatness"
symbol to start the related calibration. See "Power Cal dialog – Ref. Receiver"
on page 599, "Power Cal dialog – Meas. Receiver" on page 600, "Power Cal dialog
– Source Flatness" on page 604, and "Power Cal dialog – Leveling Table"
on page 602.
A green checkmark
indicates that the calibration data has been acquired success-
fully. A half circle overlay indicates that only the respective port frequencies are cov-
ered (see "Calibrate Only Port Frequency" on page 600).
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
Cal softtool