Command reference
R&S
®
ZNA
1568User Manual 1178.6462.02 ─ 29
Parameters:
<Source> Identifier for the source of the calibrated wave:
'A<n>' or 'PORT<n>' denote correction data for the analyzer port
no. <n>.
'ESRC<n>' or 'GEN<n>' denote correction data for the external
generator no. <n>.
<CorrData> Power correction values either in ASCII or block data format,
depending on the current FORMat[:DATA] setting.
All numbers are interpreted as dB-values.
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See "Apply" on page 599
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>? <SettingType>
Gets settings of the available power calibrations.
Suffix:
<Ch>
.
Number of the calibrated channel.
<Wv> Number of the power calibration.
Use SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:
PARameter<Wv>:COUNt? on page 1569 to query the number
of available power calibrations.
Query parameters:
SettingType WAVE | STARt | STOP | POINts | STYPe | ATTenuation |
CPOWer | CFRequency | TSTamp | LTSTamp | FSMode
WAVE
Calibrated port and calibration type (Gen=Source Power,
Rec=Measurement Receiver)
STARt
Start Frequency or Power
Stop Frequency or Power
POINts
Nr. of sweep points
STYPe
Sweep type or grid (LIN, LOG, SEGM)