7.3.19.2 TRIGger:CHANnel:AUXiliary...
The TRIGger:CHANnel<Ch>:AUXiliary... commands control the external trigger
outputs Trigger Out A to D on the Trigger board (hardware option R&S ZNA-B91).
Each trigger output can generate:
TTL pulses of configurable duration and polarity which can be sent before or after
each sweep, sweep segment, partial measurement or measurement point.
TTL high/low signals indicating an active sweep, segment, partial measurement or
measurement point.
TRIGger:CHANnel<Ch>:AUXiliary<n>:DELay................................................................. 1659
TRIGger:CHANnel<Ch>:AUXiliary<n>:DURation.............................................................1659
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval...............................................................1660
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition.............................................................. 1661
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe.................................................................1661
TRIGger:CHANnel<Ch>:AUXiliary<n>:DELay <TrigOutDelay>
Specifies the delay of the trigger out signal relative to the selected sweep interval (see
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval).
If pulses are used as trigger out signals (see TRIGger:CHANnel<Ch>:
AUXiliary<n>:STYPe), the delay is defined relative to the start or end of the sweep
event (see TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition).
Suffix:
<Ch>
.
Channel number
<n> 1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<TrigOutDelay> Range: 0 to 30 s
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
on page 1660.
Options: R&S ZNA-B91
Manual operation: See "Delay" on page 564
TRIGger:CHANnel<Ch>:AUXiliary<n>:DURation <TrigOutDuration>
If pulses are used as trigger out signals (see TRIGger:CHANnel<Ch>:
AUXiliary<n>:STYPe), this setting specifies the width of the output trigger pulses at
Trigger Out A to D.
The trigger duration must be shorter than the selected sweep interval (see TRIGger:
CHANnel<Ch>:AUXiliary<n>:INTerval).
If the trigger duration is too long, you can modify one of the following settings:
Reduce the trigger duration, if possible.