98User Manual 1178.6462.02 ─ 29
Cable sets (optional)
The required RF and LO cables listed in Table 3-9 are not part of the standard delivery.
However, it is recommended to order the suitable cable set for 2 or 4 external test sets
together with your R&S ZNA67EXT VNA system.
Currently, Rohde & Schwarz offers cable sets R&S ZN-ZCASGO manufactured by
GORE
®
:
Variant 02, order no. 1352.1659.02 for 2 external test sets
Variant 04, order no. 1352.1659.04 for 4 external test sets
Table 3-10: Cable set R&S
ZN-ZCASGO
Connection Cable type # in Var 02 # in Var 04
R&S ZNA67 port p Source Out to
diplexer p RF IN
Diplexer p REF OUT to R&S ZNA67
port p Ref In
Diplexer p MEAS OUT to R&S ZNA67
port p Meas In
R&S ZV-Z196, 67 GHz, order no.
1306.4736.00
1.85 mm (m) / 1.85 mm (m), 61 cm
x 6
p = 1, 2
R&S ZV-Z196, 67 GHz, order no.
1306.4807.00
1.85 mm (m) / 1.85 mm (m), 91 cm
x 6
p = 3, 4 (outer test
sets)
R&S ZNA67 Conv LO Out to LO power
divider
R&S ZV-Z195 26.5 GHz
2.92 mm (m) / 2.92 mm (m), 91 cm
x 1
LO power divider (+ Adapters R&S ZV-
Z1218) to converter p LO IN
R&S ZV-Z193, 26.5 GHz
3.5 mm (m) / 3.5 mm (m), 152 cm
x 2
p = 1, 2
x 4
p = 1, ..., 4
Tightening RF cables
Tightening RF cables too strongly, can damage cables and connectors. Loose tighten-
ing can result in inaccurate measurement results.
Therefore always use an appropriate torque wrench, suitable for the type of connector.
Rohde & Schwarz offers an optional 5/16” torque wrench that fits for SMA, 3.5 mm,
2.92 mm and 1.85 mm connectors (R&S ZN-ZTW variant 35). Similar wrenches are
available for other sizes of spanner, too. For ordering information, see the R&S ZN-
ZTW data sheet or product brochure.
3.5.6.3 Diplexer control
Switchover between low frequency and high frequency mode is automatically control-
led from the R&S ZNA67EXT network analyzer unit. Use the "H/L Switch" cable to con-
nect the User Port connector on the rear panel of the analyzer to the H/L SWITCH con-
nectors on the top side of the diplexers.
Getting started with R&S