GUI reference
R&S
®
ZNA
357User Manual 1178.6462.02 ─ 29
Remote command:
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent
Connector
Allows you to specify the connector type of the related physical port.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
Use Default
Allows you to toggle between default and renormalized port impedance (or impedan-
ces) for the selected logical port and connector type.
Remote command:
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
Renormalization According to Theory of
Selects the waveguide circuit theory for renormalization. The conversion formulas of
these theories only differ if the reference impedance of at least one test port has a non-
zero imaginary part.
Refer to Chapter 4.3.2, "Reference impedances", on page 152.
Remote command:
CALCulate<Chn>:TRANsform:IMPedance:RNORmal
5.2.3 Wave tab
Selects wave quantities as measured quantities.
Background information
Refer to Chapter 4.3.5, "Wave quantities and ratios", on page 158.
Meas softtool