Command reference
R&S
®
ZNA
1542User Manual 1178.6462.02 ─ 29
SOURce:GROup:SIMultaneous:FOFFset:MOFFSet:
DVALue 0 MHz
Set minimum frequency offset for frequency shifted parallel mea-
surement to 0 MHz.
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:MODE
BANDwidth
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:
BWFactor 100
Set the minimum frequency offset for frequency shifted parallel
measurement to at least 100 times the measurement bandwidth
Manual operation: See "Parallel Measurement with Frequency Offset" on page 685
7.3.15.6 SOURce:POWer:ALC...
The following commands set up the automatic level control (ALC).
SOURce<Ch>:POWer:ALC:AUBW................................................................................ 1542
SOURce<Ch>:POWer:ALC:BANDwidth......................................................................... 1543
SOURce<Ch>:POWer<Pt>:ALC:CLAMp........................................................................ 1543
SOURce<Ch>:POWer<Pt>:ALC:COUPle....................................................................... 1544
SOURce<Ch>:POWer:ALC:CSTate............................................................................... 1544
SOURce<Ch>:POWer<Pt>:ALC:OPTimize..................................................................... 1544
SOURce<Ch>:POWer<Pt>:ALC:PIParameter.................................................................1545
SOURce<Ch>:POWer:ALC:PIParameter:ASETtling........................................................ 1545
SOURce<Ch>:POWer:ALC:PIParameter:CTIMe............................................................. 1546
SOURce<Ch>:POWer<Pt>:ALC:PIParameter:GAIN........................................................ 1546
SOURce<Ch>:POWer<Pt>:ALC:PIParameter:ITIMe........................................................1546
SOURce<Ch>:POWer<Pt>:ALC:RANGe........................................................................1547
SOURce<Ch>:POWer<Pt>:ALC:RECeiver..................................................................... 1547
SOURce<Ch>:POWer<Pt>:ALC:SOFFset...................................................................... 1547
SOURce<Ch>:POWer<Pt>:ALC[:STATe]........................................................................ 1548
SOURce<Ch>:POWer:ALC:STOLerance........................................................................1548
SOURce<Ch>:POWer<Pt>:ALC:TPOWer.......................................................................1549
SOURce<Ch>:POWer<Pt>:ALC:TPOWer:DEFault.......................................................... 1549
SOURce<Ch>:POWer<Pt>:ALC:VNETwork....................................................................1549
SOURce<Ch>:POWer:ALC:AUBW <Boolean>
Enables or disables automatic IF bandwidth setting for the ALC path in channel <Ch>.
Set SOURce<Ch>:POWer:ALC:AUBW to OFF and use SOURce<Ch>:POWer:ALC:
BANDwidth to set the bandwidth manually.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Enables automatic bandwidth setting for ALC in channel <Ch>.
OFF (0)
Disables automatic bandwidth setting for ALC in channel <Ch>.
SCPI command reference