Command reference
R&S
®
ZNA
1544User Manual 1178.6462.02 ─ 29
Manual operation: See "Clamp" on page 530
SOURce<Ch>:POWer<Pt>:ALC:COUPle <Boolean>
Enforces identical ALC settings for the individual ports in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number. This suffix is ignored; the setting applies to all
ports.
Parameters:
<Boolean> ON (1)
Enforces identical ALC settings
OFF (0)
Enables port-specific ALC settings
Manual operation: See "Couple Port Settings" on page 529
SOURce<Ch>:POWer:ALC:CSTate <Boolean>
Enables or disables Automatic Level Control (ALC) for channel <Ch>.
In addition to the channel level, ALC can also be enabled or disabled at the port level
(SOURce<Ch>:POWer<Pt>:ALC[:STATe] ON).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Enable ALC for channel <Ch>
OFF (0)
Disable ALC for channel <Ch>
Manual operation: See "ALC On" on page 526
SOURce<Ch>:POWer<Pt>:ALC:OPTimize <Alc optimize>
Enables/configures ALC optimization for port <Pt> in channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Alc optimize> AUTO | ON | OFF
AUTO
The firmware decides whether speed optimization shall be per-
formed
SCPI command reference