GUI reference
R&S
®
ZNA
530User Manual 1178.6462.02 ─ 29
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:COUPle
Apply Wave Correction
See "Wave Correction" on page 359.
Remote command:
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
Settling Tolerance
Defines the maximum variation of the source signal level after the ALC has settled.
The "Settling Tolerance" has an impact on the ALC Path IF Bandwidth in "Auto" mode:
other things being equal, the smaller the "Settling Tolerance" the smaller the auto-
selected filter bandwidth.
Remote command:
SOURce<Ch>:POWer:ALC:STOLerance
Show Advanced Settings
Shows or hides advanced ALC settings:
ALC Optimize ← Show Advanced Settings
Enables/configures ALC optimization.
"Auto" (default)
The firmware decides whether speed optimization is performed.
"On"
Optimization active
"Off"
Optimization inactive
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:OPTimize
Clamp ← Show Advanced Settings
Suspends the ALC mechanism while the analyzer acquires measurement data. The
automatic level adjustment during the measurement is controlled via "Clamp" and Con-
trol Range +/-.
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:CLAMp
Stimulus softtool