Command reference
R&S
®
ZNA
1410User Manual 1178.6462.02 ─ 29
*RST: 0
Manual operation: See "Offset > Calculate after De-/Embed." on page 753
[SENSe<Ch>:]CORRection:EWAVe[:STATe] <Boolean>
Activates or deactivates the wave correction logic for wave quantities and ratios.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> *RST: ON (1)
Manual operation: See "Wave Correction" on page 359
[SENSe:]CORRection:IMEThod
<InterpolationMethod>
Selects the algorithm for the interpolation of user system error corrections.
For the factory system error correction data, the analyzer always uses linear interpola-
tion.
Parameters:
<InterpolationMethod>LINear | HORDer
LINear
Linear interpolation
HORDer
Higher order (cubic spline) interpolation
Manual operation: See "Interpolation" on page 860
[SENSe<Ch>:]CORRection:LOSS<PhyPt> <LossAtFrequency>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond <LossAtFrequency>
These commands define the offset loss at the reference frequencies f
1
and f
2
.
The reference frequencies can be set using [SENSe<Ch>:]CORRection:
LOSS<PhyPt>:FREQuency and [SENSe<Ch>:]CORRection:LOSS<PhyPt>:
SECond:FREQuency.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<LossAtFrequency> Loss at reference frequency
Range: -200 dB to +200 dB
Increment: 0.001 dB
*RST: 0 dB
Default unit: dB
SCPI command reference