GUI reference
R&S
®
ZNA
416User Manual 1178.6462.02 ─ 29
This dialog offers the same settings as the Scalar Mixer Meas setup dialog. However,
for vector mixer measurements the LO ports must be controlled by the analyzer.
5.2.15 Two Tone Group Dly tab (frequency-converting DUT)
Sets up the channel for a two-tone group delay measurement (a.k.a. mixer group delay
measurement with embedded LO), and selects the measured quantities.
These measurements require software option R&S ZNA-K9. If this option is not instal-
led, the "Two Tone Group Dly" tab is not available.
Background information
Refer to Chapter 4.7.3.6, "Embedded LO mixer group delay measurements",
on page 274.
5.2.15.1 Controls on the Two Tone Group Dly tab
If external switch matrices are used, "Two Tone Group Dly" are not supported and
hence the controls on this tab are disabled.
Meas softtool