Command reference
R&S
®
ZNA
1444User Manual 1178.6462.02 ─ 29
If the stop frequency entered is smaller than the current start frequency, the start fre-
quency is set to the stop frequency minus the minimum frequency span.
7.3.14.12 [SENSe:]GDEVice
Configures generic device commands for the related channel.
SENSe:GDEVice:ADDRess?.........................................................................................1444
[SENSe<Ch>:]GDEVice:DELete....................................................................................1444
[SENSe<Ch>:]GDEVice:MSELect..................................................................................1444
[SENSe<Ch>:]GDEVice:SELect.................................................................................... 1445
SENSe:GDEVice:ADDRess?
Returns the addresses of the configured generic external devices.
Return values:
<Address> Comma-separated list of string values.
These addresses are used for setting and unsetting the com-
mand to be executed via [SENSe<Ch>:]GDEVice:SELect
and [SENSe<Ch>:]GDEVice:DELete, respectively.
Example: See [SENSe<Ch>:]GDEVice:SELect on page 1445.
Usage: Query only
[SENSe<Ch>:]GDEVice:DELete [<Address>]
Deletes generic device command configurations for channel <Ch>.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Address> Address string
If present, only the command configuration of the related device
is deleted. Otherwise, the command configurations of all generic
external devices are deleted.
Use SENSe:GDEVice:ADDRess? to get the address strings of
the configured generic devices.
Example: See [SENSe<Ch>:]GDEVice:SELect on page 1445.
Usage: Setting only
Manual operation: See "Reset" on page 926
[SENSe<Ch>:]GDEVice:MSELect <Address>, <CommandFilepath>,
<MixerCommandId>
Defines or queries the generic device calibration mixer command configuration of
channel <Ch>.
SCPI command reference