Command reference
R&S
®
ZNA
1051User Manual 1178.6462.02 ─ 29
Before executing this command, make sure that:
●
The test coupon measurement (using CALCulate:FMODel:SFD:COUPon:
MEASure) finished successfully, or the test coupon data were successfully loaded
from file (using CALCulate:FMODel:SFD:COUPon:MEASure:FILename
●
The measurement of DUT + test fixture (using CALCulate:FMODel:SFD:DUT:
MEASure) finished successfully.
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they are
overwritten.
Usage: Event
Manual operation: See "Run <Fixture Modeling Tool>" on page 734
CALCulate:FMODel:SFD<Ph_pt>:RUN[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the deembedding files shall be assigned after the tool has run.
Suffix:
<Ph_pt>
.
Physical port number for single-ended deembedding, logical port
number for balanced deembedding.
Parameters:
<Boolean> 1 (ON) if a deembedding file shall be assinged to port <Pt>, 0
(OFF) otherwise
Manual operation: See "Apply" on page 735
CALCulate:FMODel:SFD:TOTaldiffcfg <SFDPortConfig>
Tells the SFD tool about the port ordering of the test fixture.
Parameters:
<SFDPortConfig> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Manual operation: See "Total Port Ordering" on page 741
CALCulate:FMODel... (other)
CALCulate:FMODel:DIRectory...................................................................................... 1051
CALCulate:FMODel:DIRectory:DEFault..........................................................................1052
CALCulate:FMODel:DIRectory:DEFault:CLEar................................................................1052
CALCulate:FMODel:REName........................................................................................1052
CALCulate:FMODel:DIRectory <String>
Defines the common working directory of the ISD, SFD and EZD fixture modeling tools.
SCPI command reference