GUI reference
R&S
®
ZNA
699User Manual 1178.6462.02 ─ 29
PuMo In
If external Pulse Modulator Control is enabled, this setting allows you to select
between positive or negative polarity of the "PuMo In" signal.
Remote command:
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]
Trig. Out
Allows you to enable/disable pulse modulator control via the respective "Trig. Out".
Note that a Trigger Out that has been previously configured to indicate sweeps, seg-
ments or measurement points for this channel, cannot be used for pulse generation.
The corresponding table row is disabled in this case.
To change the trigger configuration, use the "Trigger Out Manager..." button to open
the Trigger Out Manager dialog.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval PGENerator preceded by
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition BEFore); TRIGger:
CHANnel<Ch>:AUXiliary<n>:INTerval NONE
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe POSitive | NEGative
Measurement settings
Sets up the "Pulse Profile" measurement for this channel (see Chapter 5.10.3, "Sweep
Type tab", on page 546).
Acq Delay/Fixed
Defines the acquisition delay relative to the master pulse. Can only be set in synchron-
ized measurement mode. "Fixed" means that Autoset does not change the configured
value.
Remote command:
[SENSe<Ch>:]PULSe:MEASdelay
ALC Acq Delay/Fixed
Defines the acquisition delay of the ALC measurement relative to the master pulse.
Can only be set in synchronized measurement mode and if ALC optimization is not
enforced. "Fixed" means that Autoset does not change the configured value.
Remote command:
[SENSe<Ch>:]PULSe:ALC:MEASdelay