Command reference
R&S
®
ZNA
1467User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin:SEGMent<Seg>:COUNt.............................. 1470
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin:SEGMent<Seg>:STARt............................... 1471
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin:SEGMent<Seg>:STATe............................... 1471
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin:SEGMent<Seg>:STOP................................1471
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin[:STATe]...................................................... 1472
[SENSe<Ch>:]PULSe:GENerator<Id>:WIDTh.................................................................1472
[SENSe<Ch>:]PULSe:INTernal[:STATe]..........................................................................1472
[SENSe<Ch>:]PULSe:MEASdelay.................................................................................1472
[SENSe<Ch>:]PULSe:PERiod.......................................................................................1473
[SENSe<Ch>:]PULSe:PORT<Pt>:DELay....................................................................... 1473
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]..................................1473
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut:INVerted[:STATe].............................. 1474
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut[:STATe]........................................... 1474
[SENSe<Ch>:]PULSe:PORT<Pt>[:STATe]...................................................................... 1474
[SENSe<Ch>:]PULSe:PORT<Pt>:WIDTh....................................................................... 1475
[SENSe<Ch>:]PULSe[:STATe].......................................................................................1475
[SENSe<Ch>:]PULSe:SYNCron[:STATe]........................................................................ 1475
[SENSe<Ch>:]PULSe:ALC:MEASdelay <Measurement Delay>
Defines the acquisition delay of the ALC measurement relative to the master pulse.
Can only be set in synchronized measurement mode ([SENSe<Ch>:]PULSe:
SYNCron[:STATe] ON) and if SOURce<Ch>:POWer<Pt>:ALC:OPTimize is not set
to ON.
Suffix: 
<Ch>
.
Channel number
Parameters:
<Measurement
Delay>
Default unit: s
Options:  R&S ZNA-K7
Manual operation:  See "ALC Acq Delay/Fixed" on page 699
[SENSe<Ch>:]PULSe:ATIMe <Acquisition time>
Defines the duration of the data acquisition for pulse measurements in channel <Ch>.
Suffix: 
<Ch>
.
Channel number
Parameters:
<Acquisition time> Data acquisition time
Default unit: s
Manual operation:  See "Acq Time/Fixed" on page 700
[SENSe<Ch>:]PULSe:COUPled[:STATe] <Boolean>
Enforces identical pulse generator settings for all ports.
SCPI command reference