Command reference
R&S
®
ZNA
1475User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]PULSe:PORT<Pt>:WIDTh <pulse width for pulse modulators and
trigger outs>
Defines the port-specific width of the pulses generated by the respective internal pulse
generator.
This setting only takes effect, if:
Internal Pulse Control is enabled ([SENSe<Ch>:]PULSe:INTernal[:STATe]
ON)
Fixed Duty Cycle is disabled ([SENSe<Ch>:]PULSe:FXDCycle[:STATe] OFF)
Suffix:
<Ch>
.
Channel number
<Pt> VNA port number
Parameters:
<pulse width for pulse
modulators and
trigger outs>
Default unit: s
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Pulse Delay/Pulse Width" on page 697
[SENSe<Ch>:]PULSe[:STATe] <Boolean>
Activates or deactivates pulse generation.
Suffix:
<Ch>
.
Parameters:
<Boolean> ON (1)
Pulse generation activated.
OFF (0)
Pulse generation deactivated.
*RST: OFF (0)
Manual operation: See "Pulse Generator" on page 692
[SENSe<Ch>:]PULSe:SYNCron[:STATe] <Boolean>
Switches between synchronous and asynchronous pulse measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Synchronous measurement, i.e. the pulse measurement is
synchronized to the generated pulses.
SCPI command reference

Table of Contents