Command reference
R&S
®
ZNA
1472User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]PULSe:GENerator<Id>:TRAin[:STATe] <Set pulse train active/
inactive>
Suffix:
<Ch>
.
<Id>
Parameters:
<Set pulse train
active/inactive>
Manual operation: See "Enable Pulse Train" on page 697
[SENSe<Ch>:]PULSe:GENerator<Id>:WIDTh
<Width for the generated pulse>
Suffix:
<Ch>
.
<Id>
Parameters:
<Width for the
generated pulse>
Default unit: s
Manual operation: See "Pulse Delay/Pulse Width" on page 697
[SENSe<Ch>:]PULSe:INTernal[:STATe] <Boolean>
Switches between internal and external pulse control.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Internal pulse control
OFF (0)
External pulse control
*RST: ON
Options: R&S ZNA-B91
Manual operation: See "Pulse Modulator Control" on page 698
[SENSe<Ch>:]PULSe:MEASdelay <Value>
Defines the measurement delay (acquisition delay) relative to the pulse start.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> Measurement delay [s]
SCPI command reference