Command reference
R&S
®
ZNA
1517User Manual 1178.6462.02 ─ 29
Example:
*RST
Reset the analyzer, activating a linear frequency sweep.
SWE:TYPE LOG
Change to a logarithmic frequency sweep, resetting the stimulus
values of the sweep points.
Manual operation: See "Lin Freq" on page 546
7.3.14.23 [SENSe:]... (other)
[SENSe<Ch>:]COUPle................................................................................................. 1517
[SENSe<Ch>:]IFPath................................................................................................... 1517
[SENSe<Ch>:]PAMPlifier<Pt>[:STATe]........................................................................... 1518
[SENSe<Ch>:]PAMPlifier2:VALue..................................................................................1518
[SENSe<Ch>:]PHASe:MODE........................................................................................1519
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence.................................................................... 1519
[SENSe:]UDSParams:ACTive........................................................................................1520
[SENSe:]UDSParams<Pt>:PARam................................................................................ 1520
[SENSe<Ch>:]COUPle <Mode>
Determines the order of partial measurements and sweeps (see Chapter 4.1.4.1, "Par-
tial measurements and driving mode", on page 111).
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the sweep mode applies
to all channels in the active recall set.
Parameters:
<Mode> ALL | AUTO | NONE
AUTO – automatic mode: fast sweeps are performed in alterna-
ted driving mode, slower sweeps in chopped mode
ALL – chopped driving mode, complete all partial measurements
before proceeding to the next sweep point
NONE – alternated driving mode on, reverse the order of partial
measurements and sweeps
*RST: NONE
Example:
COUP NONE
Activate the alternated driving mode.
TRIG:LINK 'PPO'
Set the triggered measurement sequence equal to one partial
measurement. Each trigger event starts one partial measure-
ment for all sweep points.
Manual operation: See "Driving Mode" on page 687
[SENSe<Ch>:]IFPath <If Path>
Selects the analog IF signal path.
SCPI command reference