Command reference
R&S
®
ZNA
1421User Manual 1178.6462.02 ─ 29
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation: See "Apply" on page 578
[SENSe<Chn>:]CORRection:PSTate?
Gets the power calibration label of the active trace in channel <Chn>.
See Chapter 4.5.6.3, "Power calibration labels", on page 222 for the possible return
values.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Query only
Manual operation: See "Apply" on page 599
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel <Chn>.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see Chapter 4.5.4, "Calibration labels", on page 206).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See [SENSe<Ch>:]CORRection:DATE?
Usage: Query only
Manual operation: See "Apply" on page 578
7.3.14.11 [SENSe:]FREQuency...
The [SENSe:]... commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
SCPI command reference