Command reference
R&S
®
ZNA
1420User Manual 1178.6462.02 ─ 29
Same command structure and logic as for [SENSe<Ch>:]CORRection:CDATa...
commands, except for additional parameters SourceVNAPort and LoadVNAPort
that restrict the returned correction data to the measurement path between by the
respective physical VNA ports (if any). See Chapter 4.7.41.5, "Multiple paths and cali-
bration", on page 329 for details.
Suffix:
<Ch>
.
<PhyPt>
Parameters:
<ErrorTerm>
<SourceTestPort>
<LoadTestPort>
<SourceVNAPort> Number of the VNA port connected to the source test port (via
switch matrix)
<LoadVNAPort> Number of the VNA port connected to the load test port (via
switch matrix)
<CorrectionData> Correction data either in ASCII or block data format, depending
on the current FORMat[:DATA] setting.
[SENSe<Ch>:]CORRection[:STATe] <Boolean>
Enables or disables the system error correction for channel <Ch>.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the correction.
*RST: ON
Example:
*RST; :CORR?
Reset the instrument and query whether channel 1 is system
error corrected. The response is 1.
Manual operation: See "User Cal Active" on page 635
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt> Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S ZNA-K4).
SCPI command reference