Command reference
R&S
®
ZNA
1419User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Calibrated channel number
Example: See SOURce<Ch>:POWer:CORRection:MIXer:RF[:
ACQuire] on page 1664.
Usage: Event
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe] <Boolean>
Enables or disables the receiver power calibration for channel <Ch> and for the
received waves b
<PhyPt>
. The setting command is disabled unless the received waves
have been power calibrated ([SENSe<Ch>:]CORRection:POWer<PhyPt>:
ACQuire BWAVe, ...). The query always returns a result.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the receiver power calibration
for the received waves b
<PhyPt>
.
*RST: OFF
Example:
*RST; CORR:POW:ACQ BWAVe,1,PORT,2
Perform a receiver power calibration of the wave b1 using port 2
as a source port, assuming that the source power a
2
is correct.
A through connection from port 2 to port 1 is needed.
CORR:POW?
Check whether the calibration is applied (the response is 1).
CALC:PAR:MEAS 'TRC1', 'B1D2'
Select b
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no. <Mk> in the center of the sweep range and
query the measurement value. The calibrated power of the
received wave b
1
is approx. equal to the default source power
value.
See also example for [SENSe<Ch>:]CORRection:
POWer<PhyPt>:ACQuire.
Manual operation:
See "Port Overview" on page 637
[SENSe<Ch>:]CORRection:SMATrix:CDATa <ErrorTerm>, <SourceTestPort>,
<LoadTestPort>, <SourceVNAPort>, <LoadVNAPort>, <CorrectionData>...
[SENSe<Ch>:]CORRection:SMATrix:CDATa:PORT<PhyPt> <ErrorTerm>,
<SourceTestPort>, <LoadTestPort>, <SourceVNAPort>, <LoadVNAPort>,
<CorrectionData>...
Writes or reads system error correction data in the presence of switching matrices.
SCPI command reference