Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
HARMonic[:ACQuire] on page 1571.
Usage: Event
Manual operation: See "Harmonic Orders to Calibrate" on page 590
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note: The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire is enabled
only after steps 1 and 2 have been carried out (SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:LTONe[:ACQuire], SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:UTONe[:ACQuire]).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
([SENSe<Ch>:]FREQuency:IMODulation:RECeiver).
Setting parameters:
<CalOnlyPortFreq> OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST: 0
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
IMODulation:LTONe[:ACQuire]
Usage: Setting only
[SENSe<Ch>:]CORRection:POWer:MIXer:IF:ACQuire
Starts the IF receiver calibration (2nd power calibration step for scalar mixer measure-
ments), stores and applies the calibration data.
Note: The receiver calibration relies on the IF source power calibration acquired in
step no. 1. [SENSe<Ch>:]CORRection:POWer<PhyPt>:MIXer:IF:ACQuire is
enabled only after step 1 has been carried out
(SOURce<Ch>:POWer<PhyPt>:CORRection:MIXer:RF[:ACQuire]).