Command reference
R&S
®
ZNA
1572User Manual 1178.6462.02 ─ 29
Example:
*RST; FREQ:STAR 0.5E+9; STOP 1E+9
Reset the analyzer (activating a frequency sweep) and set the
sweep range between 1 GHz and 1.1 GHz.
CALCulate:PARameter:MEASure '<TraceName>',
'h3b2(p1)'
Configure a 3rd harmonic measurement with source port 1 and
receive port 2.
SOUR:POW:CORR:PMET:ID 1
Select power meter no. 1 (previously configured in the External
Power Meters dialog and properly connected) for the RF source
calibration.
SOUR:POW1:CORR:HARM
Start source power calibration at source port 1.
CORR:POW:HARM:ACQ
Start receiver power calibration at receive port 2.
Usage:
Event
Options: R&S ZNA-K4
Manual operation: See "Harmonic Orders to Calibrate" on page 590
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
Starts the source calibration for the lower tone (1
st
power calibration step for intermo-
dulation measurements), stores and applies the calibration data. The external power
meter used is selected via SOURce<Ch>:POWer<PhyPt>:CORRection:PMETer:ID.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Port number This suffix is ignored; the analyzer calibrates the
port that is selected as a source port for the intermodulation
measurement
([SENSe<Ch>:]FREQuency:IMODulation:LTONe).
SCPI command reference