Command reference
R&S
®
ZNA
1476User Manual 1178.6462.02 ─ 29
OFF (0)
Asynchronous measurement
*RST: ON (1)
Options: R&S ZNA-K7
Manual operation: See "Sync Meas to Pulse Gen." on page 700
7.3.14.19 [SENSe:]ROSCillator...
The [SENSe:]ROSCillator... commands control the frequency reference signal.
[SENSe:]ROSCillator:EXTernal:FREQuency................................................................... 1476
[SENSe:]ROSCillator:SMA:INPut...................................................................................1476
[SENSe:]ROSCillator:SMA:OUTPut............................................................................... 1477
[SENSe<Ch>:]ROSCillator[:SOURce]............................................................................ 1477
[SENSe:]ROSCillator:EXTernal:FREQuency <ExtClockFreq>
Specifies or queries the frequency of the external reference oscillator.
Parameters:
<ExtClockFreq> Frequency of the external reference clock signal.
Range: See the data sheet of your analyzer.
*RST: 10 MHz
Default unit: Hz
Example:
ROSC EXT
Select the external reference clock as clock source.
ROSC:EXT:FREQ 10MHz
Specify the frequency of the external reference clock.
ROSC:EXT:FREQ?
Query the frequency of the external reference oscillator. The
response is 10000000 Hz.
Manual operation: See "Ext Frequency" on page 887
[SENSe:]ROSCillator:SMA:INPut <SupportedFreqs>
Specifies the frequency of the incoming reference clock signal at the Reference In/Out
SMA connector (see [SENSe<Ch>:]ROSCillator[:SOURce]).
Parameters:
<SupportedFreqs> F100mhz | F1GHz
100 MHz or 1 GHz
*RST: F100mhz
Manual operation: See "Reference I/O (SMA) > In" on page 887