GUI reference
R&S
®
ZNA
354User Manual 1178.6462.02 ─ 29
For True Differential Mode and Source Coherence... see their description in the "Pre-
defined Config" tab.
Define Physical to Logical Port Relation
Allows you to define balanced, single-ended, and unused ports and provides functions
for (re)numbering the resulting logical ports.
In principle, it is possible to combine any pair of two physical analyzer ports into a bal-
anced port. With n test ports a maximum of n/2 (n even) or (n – 1)/2 (n odd) logical
ports is supported.
To define a balanced port, select two physical ports and tap "Balanced".
To dissolve balanced ports, select them and tap "Single".
To exclude logical ports from the measurement, select them and tap "Unused".
To assign a number to a logical port, select it and enter a new "Logical Port" num-
ber in the corresponding field
To number the logical ports in line with the physical ports, select "Identity 1-1"
For balanced ports, the lower of the two physical port numbers is selected.
To number the logical ports consecutively from top to bottom, select "Aligned
1,2,3..."
Remote command:
SOURce<Ch>:LPORt<LogPt>
SOURce<Ch>:LPORt<LogPt>:CLEar
User Config tab
The "User Config" tab of the "Balanced Ports" dialog defines a new balanced port con-
figuration.
If a switch matrix is configured, this tab is replaced by the Logical Port Config tab tab.
Meas softtool