Command reference
R&S
®
ZNA
1532User Manual 1178.6462.02 ─ 29
*RST: OFF
Manual operation: See "Independent Gen" on page 921
7.3.15.4 SOURce:RLO...
The SOURce:RLO commands configure the optional rear LO Out port (R&S ZNA-B8).
SOURce<Ch>:RLO:FREQuency................................................................................... 1532
SOURce<Ch>:RLO:PABSolut....................................................................................... 1533
SOURce<Ch>:RLO:PERMenable.................................................................................. 1533
SOURce<Ch>:RLO:POFFset........................................................................................ 1533
SOURce<Ch>:RLO:FREQuency <Numerator>, <Denominator>, <Offset>,
<SweepType>
Defines the frequency formula at the (optional) rear LO Out port for channel <Ch>.
Note: The frequency formula is applied even if the analyzer returns an error message,
because the frequency is outside the allowed range.
Suffix:
<Ch>
.
Channel number
Parameters:
<Numerator>,
<Denominator>,
<Offset>
Parameters of the port frequency formula
f
p
= <Numerator>/<Denominator> * f
b
+ <Offset> .
For <SweepType> FB the variable f
b
represents the channel
base frequency, for <SweepType> CW it is set to 0 Hz.
Note: The <Offset> parameter also includes the "Offset Ratio" in
manual control.
Range: The converted frequency or frequency range must
be within the analyzer's frequency range, depend-
ing on the instrument model. In addition the
<Numerator> values are rounded to positive or neg-
ative integer numbers; zero is not allowed.
*RST: The default frequency/frequency range corresponds
to the sweep range or CW frequency of the ana-
lyzer, i.e. <Numerator> = <Denominator> = 1, <Off-
set> = 0 Hz.
<SweepType> FB | CW
FB
Sweep with channel base frequency f
b
CW
Fixed frequency
Options: R&S ZNA-B8
Manual operation: See "Freq. Conversion" on page 672
SCPI command reference