Start Frequency/Stop Frequency/Number of Points/Freq Step Size/Base Power
Standard parameters of a linear frequency sweep (see Chapter 5.8, "Stimulus softtool",
on page 522).
Start Power/Stop Power/Number of Points/Fixed CW
Standard parameters of a power sweep (see Chapter 5.8, "Stimulus softtool",
on page 522).
Delta Frequency
Defines the frequency difference between the upper and the lower tone for linear fre-
quency and power sweeps. The tone distance is kept constant across the entire
sweep.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:TDIStance
Start delta Freq./Stop delta Freq./Number of Points/Base Power/CW Fre-
quency/CW fixed to
Defines the lower tone and upper tone frequency sweeps for IMD Sweep Type "Delta
F".
Additional channel settings
IF Bandwidth
See "Bandwidth" on page 536.
Step Att Receiving Port
Only available if the selected Receiving Port is equipped with a receiver step attenua-
tor (see Chapter 4.7.31, "Receiver step attenuators", on page 309).
Defines the receiver step attenuation at this port.
Frequency Overview dialog
The "Frequency Overview" dialog summarizes the resulting measured frequency
ranges, according to the current setup of the related channel. Invalid frequency ranges
(that are not covered by the analyzer's frequency range) are marked in red.
Access: "Frequency Overview" in various channel mode setup dialogs