EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Logical Port Config tab

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
353User Manual 1178.6462.02 ─ 29
The port configurations are arranged in the list to the right. The resulting port assign-
ment is shown on the left-hand side of the "Predefined Config" tab.
For a single-ended port, the diagram shows a single line between the physical test
port and the logical port.
For a balanced port, two physical ports are combined to form a single logical port.
For unused ports, the physical port is crossed out; no logical port number is
assigned.
Select Predefined Port
Allows you to select from a set of predefined logical port configurations.
The available configurations depend on the number of analyzer ports.
Remote command:
SOURce<Ch>:LPORt<LogPt>
SOURce<Ch>:LPORt<LogPt>:CLEar
True Differential Mode
Activates/deactivates True Differential Mode
This checkbox is only visible if True differential mode option R&S ZNA-K61 is available.
It is enabled, if at least one balanced port is configured, whose physical ports are fed
by independent sources (see Chapter 4.7.20, "Internal 3rd and 4th source for 4-port
R&S ZNA", on page 302). In case the balanced port is made up of frequency converter
ports, Power Control must be active for both of them.
Use the "Source Cohrence..." button to open the Source Coherence dialog dialog,
which allows you to configure the TruDi-specific sweep types.
Remote command:
SOURce<Ch>:TDIF[:STATe]
Source Coherence...
Opens the dialog of the same name, which allows you to configure and enable source
coherence and to configure the true differential mode sweep types "Phase Imbalance"
and "Amplitude Imbalance".
See Chapter 5.13.3.2, "Source Coherence dialog", on page 703.
Logical Port Config tab
The "Logical Port Config" tab of the "Balanced Ports" dialog allows you to configure
logical ports (balanced or unbalanced) and to disable unused physical ports in the
active channel.
This tab is only visible if a switch matrix is configured).
Meas softtool

Table of Contents

Related product manuals