GUI reference
R&S
®
ZNA
667User Manual 1178.6462.02 ─ 29
Wait (Sweep Sequencer Table)
The cells in the "Wait" column define the delay times between subsequent sequence
steps and between the final sequence step and the sweep (segment) start.
Remote command:
Unsegmented sweeps
CONTrol<Ch>:SEQuence<Nr>:DELay
Segmented sweeps:
CONTrol<Ch>:SEGMent<SegNr>:SEQuence<SeqNr>:DELay
5.12.2 Port Config tab
Configures the source and receive ports of the R&S ZNA, for measurements on fre-
quency-converting DUTs, defining arbitrary port frequencies and powers. While the
arbitrary mode is active, "Arb" and the selected stimulus axis are displayed in the chan-
nel line.
Background information
Refer to Chapter 4.7.3, "Frequency conversion measurements", on page 261.
5.12.2.1 Controls on the Port Config tab
If either multiple DUTs are configured (see Chapter 5.12.2.3, "Define Parallel Measure-
ment dialog", on page 682) or the "Fixture Simulator" is disabled for the related chan-
nel (see "Fixture Simulator" on page 655), the "Balanced Ports..." button is inactive
(grayed out).
Port Settings...
Opens a dialog that allows you to configure the source and receiver ports of the
R&S ZNA for arbitrary frequencies, source and receiver levels: see Chapter 5.12.2.2,
"Port Settings dialog", on page 668.
Combiner
Tells the analyzer if and how signals from two ports are combined. For descriptions and
visualizations of the supported combiner configurations, see "Combiner Configuration"
on page 383.
Channel Config softtool