4.7.20 Internal 3rd and 4th source for 4-port R&S ZNA
R&S ZNAxx-B3
A four-port R&S ZNA has at least two internal sources. With two internal sources
(default), the physical ports of a four-port VNA are split into groups P1 = {1,2} and P2 =
{3,4} such that source 1 can only drive ports in P1 and source 2 can only drive ports in
P2. Independent signals require the source ports to reside in different port groups. For
example, you cannot use VNA ports 1 and 2 as independent source ports.
Hardware option R&S ZNAxx-B3 expands a four-port network analyzer R&S ZNAxx
with a third and fourth internal source. With four independent internal sources, a wide
range of Frequency conversion measurements can be performed, without the need for
additional (external) generators.
Optional extensions and accessories