Concepts and features
R&S
®
ZNA
294User Manual 1178.6462.02 ─ 29
Noise power spectral densitiy
The noise power spectral density, or noise density for short, is calculated as
NPSD = ND = N
meas, corr
/ B
where N
meas, corr
denotes the measured sideband-corrected noise power, and B the
equivalent noise bandwidth (see "Receiver noise temperature" on page 291). N
meas, corr
is basically independent of what is currently connected to the test ports, so no distinc-
tion between calibration and DUT is required.
4.7.12 Measurement uncertainty analysis
Option R&S ZNA-K50/R&S ZNA-K50P
Measurement uncertainty analysis is available if option R&S ZNA-K50 and METAS
VNA Tools (version 2.2.6 or higher) are installed on the instrument.
Option R&S ZNA-K50P offers the same functionality as option R&S ZNA-K50, but
comes with METAS VNA Tools preinstalled. It can be shipped with new instruments or
installed at Rohde&Schwarz service.
The purpose of the METAS VNA Tools software is to compute uncertainties of S-
parameter measurements. After registration, you can download it from the VNA Tools
Internet site (https://www.metas.ch/vnatools).
The uncertainty analysis of the METAS VNA Tools is limited to S-parameter mea-
surements. Uncertainty analysis of wave quantities is not supported.
The uncertainty calculations are in line with EURAMET Calibration Guide No. 12
Version 3.0 (03/2018).
If external switch matrices are used, measurement uncertainty analysis is not sup-
ported.
METAS Calibration
The VNA firmware directly interacts with the DLLs of the METAS VNA Tools. During a
"METAS Calibration", the VNA firmware collects all data that are required to set up a
METAS VNA Tools project. After the "METAS Calibration"METAS calibration is comple-
ted, the analyzer is ready for real-time uncertainty analysis: after each sweep it passes
the raw (uncalibrated) S-parameters to the VNA tools and receives the calibrated S-
parameters plus uncertainties (95% confidence ellipse in the complex plane) in return.
The calibrated S-parameter traces plus their uncertainty bands can be displayed in
magnitude, phase, real and imaginary trace formats. The display of uncertainties in
complex diagrams (polar, Smith) is not supported.