GUI reference
R&S
®
ZNA
361User Manual 1178.6462.02 ─ 29
To configure external generators, use the "External Generators" dialog. See Chap-
ter 4.7.40, "External generators", on page 321.
Properties
Allows you to select the measurement detector and the display unit for the selected
wave quantity.
Detector ← Properties
Selects the algorithm that is used to calculate the results points from the raw measure-
ment data.
For details refer to Chapter 4.3.5.3, "Detector settings", on page 160.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
[SENSe<Ch>:]SWEep:DETector:TIME
Display unit ← Properties
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or to convert it into a power value P according to the formula
P = V
2
/ Re(Z
0
).
Z
0
denotes the reference impedance of the source port (for wave quantities a
n
) or of
the receive port (for wave quantities b
n
). The reference impedances are defined in the
"Balanced Ports" dialog; see Chapter 5.2.2.5, "Balanced Ports dialog", on page 352.
Remote command:
CALCulate<Chn>:FORMat:WQUType
5.2.4 Ratio | Harmonics tab
The "Ratio | Harmonics" tab gives access to two types of measured quantities:
Ratios of wave quantities
Harmonics, i.e. arbitrary harmonics, ratios between harmonics and the fundamental
wave, and total harmonic distortion (THD)
To this end, the source remains at the fundamental frequency whereas the receiver
is set to n times the fundamental frequency.
Access: Trace – [Meas] > "S-Params" > "Harmoniocs"
Meas softtool