EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - 5.19.7.4 Switch Matrix RF Connections dialog

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
914User Manual 1178.6462.02 ─ 29
simply use "Copy from local network" to copy the "Subnet Mask" and "Standard Gate-
way" from the analyzer's IP settings.
5.19.7.4 Switch Matrix RF Connections dialog
The "Switch Matrix RF Connections" dialog allows you to define the RF connections
between the VNA and the switch matrices, and the test port numbering.
Access: System – [Setup] > "External Ports" > "RF Connections..."
Background information
Refer to section Chapter 4.7.41, "External switch matrices", on page 323.
Defining switch matrix RF connections typically involves the following steps.
1. Add one or more external switch matrices to the RF connection configuration.
These matrices must have been registered before; see Chapter 5.19.7.2, "External
Matrices dialog", on page 909 for details.
2. For each matrix VNA port (the "north" ports in the graphical representation), specify
the connected VNA port or define it as "Unused" if no such connection exists.
For every matrix, there must be at least one RF connection to the VNA.
3. Select Renumber Test Ports to load the default test port assignment.
Setup softtool

Table of Contents

Related product manuals